Diode laser overheating in Gliwice — emergency water filter replacement
808 nm diode laser
An urgent on-site visit to Gliwice for an overheating diode laser. A cracked fitting near the water filter was leaking coolant. We replaced the filter and refilled the system with fresh coolant.
Problem
We received an urgent request from Gliwice about an overheating diode laser used for hair removal. An on-site technician was needed.
Diagnosis
During the cooling system inspection, our technician identified the cause. A fitting near the water filter had cracked at the connection between the filter and the cooling hose.
Coolant was leaking through the damaged connection, disrupting normal cooling of the equipment.
Work performed
During the emergency service visit, we replaced the damaged water filter and refilled the system with fresh coolant.
Result
The laser is working again. This failure highlights the importance of regularly checking filters and cooling system connections for leaks.
Laser Tech Service provides laser servicing, refurbishment and aesthetic laser repairs, including on-site visits.