Dismantling
Dismantling the device to access electronics and connections.
Cavi Shape Advanced
Comprehensive diagnostics and repair of a Cavi Shape Advanced in Wrocław. The device had several independent problems: three RF handpieces were not working, infrared light was unavailable and ultrasound was operating incorrectly.
Equipment Cavi Shape Advanced required comprehensive diagnostics of several functions at once.
Three RF handpieces were not working, the infrared light function was faulty and ultrasound was operating incorrectly.
The nature of the problems required dismantling the device and checking connections between the handpieces and electronics.
After dismantling, we checked the cables, connections, contacts and main board condition.
During diagnostics, we took electrical measurements and checked the signals responsible for individual device functions.
The measurements allowed us to check continuity and locate poor contacts.
The footage shows actual device diagnostics during servicing in Wrocław.
Open the video on YouTube →
The main board was heavily contaminated with deposits accumulated during use.
The electronics were thoroughly cleaned. We also checked tracks, contacts and connection points.
Connections requiring attention were resoldered to restore correct electrical contact.
We checked connections responsible for correct handpiece operation and signals between the handpieces and electronics.
Open the video on YouTube →The repair covered electrical connections and servicing of the device electronics.
Dismantling the device to access electronics and connections.
Thorough cleaning of contacts, connectors and the main board.
Checking tracks and repairing connections requiring soldering.
Checking RF, infrared and ultrasound after repair.
Cleaning the main board and checking connections that contributed to the device problems.
Open the video on YouTube →After contact cleaning, connection servicing and main board repair, the device was reassembled.
We restored operation of the three RF handpieces and infrared light.
We also checked ultrasound operation after servicing. The device works correctly.
After completing the work, we checked the device and repaired functions.
Open the video on YouTube →
Laser Tech Service diagnoses and repairs professional aesthetic and laser equipment.
Our work includes diagnostics of electronics, main boards, cables, connectors, RF systems, ultrasound and other treatment equipment components.
Our service centre is in Wrocław, and we also handle requests from customers throughout Poland.
Explore Laser Tech Service services →Send us the equipment name and model, photographs and a brief description of the problem. We will use this information to agree the next steps.